Measurement Specialties Inc. Model 9116 User’s Manual
Page 97 www.MEAS-SPEC.com
Following is a simple, step-by-step procedure for executing a specified-value span calibration of
a Model 9116 Intelligent Pressure Scanner. For the purposes of this example, it will be assumed
that an upscale pressure of 14.9800 psi is available from a dead weight tester for the calibration
of 15 psi internal transducers. All sixteen channels are 15 psi full scale.
Ensure that valves are in RUN/CAL mode (default)
… normal data acquisition
Perform Re-zero calibration first
Place the module calibration manifolds into the CAL position
if this is the desired pressure application input. The RUN
position may be a better choice for modules with
transducers having different ranges.
[w0C01] for CAL
[w0C00] for RUN
With a deadweight tester, apply 14.9800psi to the
appropriate module CAL and CAL REF inputs [or optionally
to RUN inputs]
Delay for settling of pneumatic inputs
Verify measured data reads near expected full scale
Instruct module to calculate new gain coefficients for all 16
channels
Place calibration manifold back into the RUN position
Store new gain coefficients to transducer non-volatile
memory
… continue normal data acquisition