4.3.2. Span Calibration Summary
Following is a simple, step-by-step procedure for executing a “full scale” span calibration of a
9116 Intelligent Pressure Scanner. It is assumed that all channels in the unit are of the same full
scale pressure range. Optional commands are shown within brackets [ ].
Ensure that valves are in RUN/CAL mode (default)
… normal data acquisition
Perform Re-zero calibration
Place the module calibration manifolds into the CAL position
if this is the desired pressure application input. The RUN
position may be a better choice for modules with
transducers having different ranges.
[w0C01] for CAL
[w0C00] for RUN
Apply exact full scale pressure to appropriate module CAL
and CAL REF inputs [or optionally to RUN inputs].
Delay for settling of pneumatic inputs
Verify that measured data reads near expected full scale
Instruct module to calculate new gain coefficients for all 16
channels
Place calibration manifold back into the RUN position
Store new gain coefficients to transducer non-volatile
memory
… continue normal data acquisition